Mattress cover and foam insert, and pressure-relieving qualities are checked for damage, fluid ingress and effectiveness:
• After each patient’s discharge OR
• On a weekly basis
Inspection Steps:
• Fully strip the mattress of all linen and pillows
• Check the outer surface of the mattress cover (top, sides, and
underside) for:
o Tears, scratches, punctures or staining/marks
o Signs of fluid penetration or contamination
• Unzip or remove mattress cover and examine:
o The inner surface of the cover for staining or damage (See
pictures in appendix B)
• The foam insert for:
o Marks, damp patches, or odours.
o Signs of degradation (e.g. crumbling, loss of shape, or firmness)
• Lift the mattress to inspect the surface in contact with the bed/trolley
frame for wear or damage
• Assess the mattress for pressure-relieving
• Check for uneven wear, dips, twists, or loss of support. (see pictures in
appendix B)
• Confirm the mattress is appropriate for the patient’s pressure ulcer risk
level (e.g. static foam vs. dynamic air)
• If unsure consult Tissue Viability Microsite for Guidance.
Actions if issues are found:
• Damaged mattress covers must be replaced
• If foam insert is visibly contaminated or degraded, the entire
mattresses must be replaced
• Nurse in charge must be informed of any faults or concerns
Nurse in charge to:
• Contact Procurement to order a replacement mattress cover or full
mattress
• Arrange for uplift and disposal of condemned mattresses
The following link includes the process and form for returning an
Air Mattress in BGH only: http://intranet/resource.asp?uid=46433
Documentation:
• The template in Appendix A must be completed to confirm that every
mattress has been checked on a weekly basis
• Any faults and actions taken must be recorded in the equipment
maintenance log.